Instrument Calibration
Every instrument has at least one input and one output. For a pressure sensor, the input would be
some fluid pressure and the output would (most likely) be an electronic signal. For a loop indicator,
the input would be a 4-20 mA current signal and the output would be a human-readable display.
For a variable-speed motor drive, the input would be an electronic signal and the output would be
electric power to the motor.
Calibration and ranging are two tasks associated with establishing an accurate correspondence
between any instrument’s input signal and its output signal. Simply defined, calibration assures the
instrument accurately senses the real-world variable it is supposed to measure or control. Simply
defined, ranging establishes the desired relationship between an instrument’s input and its output.
18.1 Calibration versus re-ranging
To calibrate an instrument means to check and adjust (if necessary) its response so the output
accurately corresponds to its input throughout a specified range. In order to do this, one must
expose the instrument to an actual input stimulus of precisely known quantity. For a pressure
gauge, indicator, or transmitter, this would mean subjecting the pressure instrument to known fluid
pressures and comparing the instrument response against those known pressure quantities. One
cannot perform a true calibration without comparing an instrument’s response to known, physical
stimuli.
To range an instrument means to set the lower and upper range values so it responds with the
desired sensitivity to changes in input. For example, a pressure transmitter set to a range of 0 to
200 PSI (0 PSI = 4 mA output ; 200 PSI = 20 mA output) could be re-ranged to respond on a scale
of 0 to 150 PSI (0 PSI = 4 mA ; 150 PSI = 20 mA).
In analog instruments, re-ranging could (usually) only be accomplished by re-calibration, since
the same adjustments were used to achieve both purposes. In digital instruments, calibration and
ranging are typically separate adjustments (i.e. it is possible to re-range a digital transmitter without
having to perform a complete recalibration), so it is important to understand the difference.
18.2 Zero and span adjustments (analog instruments)
The purpose of calibration is to ensure the input and output of an instrument reliably correspond
to one another throughout the entire range of operation. We may express this expectation in the
form of a graph, showing how the input and output of an instrument should relate. For the vast
majority of industrial instruments this graph will be linear:
This graph shows how any given percentage of input should correspond to the same percentage
of output, all the way from 0% to 100%
Things become more complicated when the input and output axes are represented by units of
measurement other than “percent.” Take for instance a pressure transmitter, a device designed to
sense a fluid pressure and output an electronic signal corresponding to that pressure. Here is a graph
for a pressure transmitter with an input range of 0 to 100 pounds per square inch (PSI) and an
electronic output signal range of 4 to 20 milliamps (mA) electric current:
Although the graph is still linear, zero pressure does not equate to zero current. This is called
a live zero, because the 0% point of measurement (0 PSI fluid pressure) corresponds to a non-zero
(“live”) electronic signal. 0 PSI pressure may be the LRV (Lower Range Value) of the transmitter’s
input, but the LRV of the transmitter’s output is 4 mA, not 0 mA.
Any linear, mathematical function may be expressed in “slope-intercept” equation form:
y = mx + b
Where,
y = Vertical position on graph
x = Horizontal position on graph
m = Slope of line
b = Point of intersection between the line and the vertical (y) axis
This instrument’s calibration is no different. If we let x represent the input pressure in units
of PSI and y represent the output current in units of milliamps, we may write an equation for this
instrument as follows:
y = 0.16x + 4
On the actual instrument (the pressure transmitter), there are two adjustments which let us
match the instrument’s behavior to the ideal equation. One adjustment is called the zero while the other is called the span. These two adjustments correspond exactly to the b and m terms of the
linear function, respectively: the “zero” adjustment shifts the instrument’s function vertically on
the graph (b), while the “span” adjustment changes the slope of the function on the graph (m). By
adjusting both zero and span, we may set the instrument for any range of measurement within the
manufacturer’s limits.
The relation of the slope-intercept line equation to an instrument’s zero and span adjustments
reveals something about how those adjustments are actually achieved in any instrument. A “zero”
adjustment is always achieved by adding or subtracting some quantity, just like the y-intercept term
b adds or subtracts to the product mx. A “span” adjustment is always achieved by multiplying or
dividing some quantity, just like the slope m forms a product with our input variable x.
Zero adjustments typically take one or more of the following forms in an instrument:
• Bias force (spring or mass force applied to a mechanism)
• Mechanical offset (adding or subtracting a certain amount of motion)
• Bias voltage (adding or subtracting a certain amount of potential)
Span adjustments typically take one of these forms:
• Fulcrum position for a lever (changing the force or motion multiplication)
• Amplifier gain (multiplying or dividing a voltage signal)
• Spring rate (changing the force per unit distance of stretch)
It should be noted that for most analog instruments, zero and span adjustments are interactive.
That is, adjusting one has an effect on the other. Specifically, changes made to the span adjustment
almost always alter the instrument’s zero point1
. An instrument with interactive zero and span
adjustments requires much more effort to accurately calibrate, as one must switch back and forth
between the lower- and upper-range points repeatedly to adjust for accuracy.
18.3 Calibration errors and testing
The efficient identification and correction of instrument calibration errors is an important function
for instrument technicians. For some technicians – particularly those working in industries where
calibration accuracy is mandated by law – the task of routine calibration consumes most of their
working time. For other technicians calibration may be an occasional task, but nevertheless
these technicians must be able to quickly diagnose calibration errors when they cause problems
in instrumented systems. This section describes common instrument calibration errors and the
procedures by which those errors may be detected and corrected.
1However, it is actually quite rare to find an instrument where a change to the zero adjustment affects the
instrument’s span.
18.3.1 Typical calibration errors
Recall that the slope-intercept form of a linear equation describes the response of any linear
instrument:
y = mx + b
Where,
y = Output
m = Span adjustment
x = Input
b = Zero adjustment
A zero shift calibration error shifts the function vertically on the graph, which is equivalent
to altering the value of b in the slope-intercept equation. This error affects all calibration points
equally, creating the same percentage of error across the entire range. Using the same example of a
pressure transmitter with 0 to 100 PSI input range and 4 to 20 mA output range:
If a transmitter suffers from a zero calibration error, that error may be corrected by carefully
moving the “zero” adjustment until the response is ideal, essentially altering the value of b in the
linear equation.
A span shift calibration error shifts the slope of the function, which is equivalent to altering
the value of m in the slope-intercept equation. This error’s effect is unequal at different points
throughout the range:
If a transmitter suffers from a span calibration error, that error may be corrected by carefully
moving the “span” adjustment until the response is ideal, essentially altering the value of m in the
linear equation.
A linearity calibration error causes the instrument’s response function to no longer be a straight
line. This type of error does not directly relate to a shift in either zero (b) or span (m) because the
slope-intercept equation only describes straight lines:
Some instruments provide means to adjust the linearity of their response, in which case this
adjustment needs to be carefully altered. The behavior of a linearity adjustment is unique to each
model of instrument, and so you must consult the manufacturer’s documentation for details on how
and why the linearity adjustment works. If an instrument does not provide a linearity adjustment,
the best you can do for this type of problem is “split the error” between high and low extremes, so
the maximum absolute error at any point in the range is minimized.
A hysteresis calibration error occurs when the instrument responds differently to an increasing
input compared to a decreasing input. The only way to detect this type of error is to do an up-down
calibration test, checking for instrument response at the same calibration points going down as going
up:

Hysteresis errors are almost always caused by mechanical friction on some moving element
(and/or a loose coupling between mechanical elements) such as bourdon tubes, bellows, diaphragms,
pivots, levers, or gear sets. Friction always acts in a direction opposite to that of relative motion,
which is why the output of an instrument with hysteresis problems always lags behind the changing
input, causing the instrument to register falsely low on a rising stimulus and falsely high on a
falling stimulus. Flexible metal strips called flexures – which are designed to serve as frictionless
pivot points in mechanical instruments – may also cause hysteresis errors if cracked or bent. Thus,
hysteresis errors cannot be remedied by simply making calibration adjustments to the instrument –
one must usually replace defective components or correct coupling problems within the instrument
mechanism.
In practice, most calibration errors are some combination of zero, span, linearity, and hysteresis
problems. An important point to remember is that with rare exceptions, zero errors always
accompany other types of errors. In other words, it is extremely rare to find an instrument with a
span, linearity, or hysteresis error that does not also exhibit a zero error. For this reason, technicians
often perform a single-point calibration test of an instrument as a qualitative indication of its
calibration health. If the instrument performs within specification at that one point, its calibration
over the entire range is probably good. Conversely, if the instrument fails to meet specification at
that one point, it definitely needs to be recalibrated.
A very common single-point test for instrument technicians to perform on differential pressure
(“DP”) instruments is to close both block valves on the three-valve manifold assembly and then
open the equalizing valve, to produce a known condition of 0 PSI differential pressure:
Most DP instrument ranges encompass 0 PSI, making this a very simple single-point check. If
the technician “blocks and equalizes” a DP instrument and it properly reads zero, its calibration is
probably good across the entire range. If the DP instrument fails to read zero during this test, it
definitely needs to be recalibrated.
18.3.2 As-found and as-left documentation
An important principle in calibration practice is to document every instrument’s calibration as it
was found and as it was left after adjustments were made. The purpose for documenting both
conditions is to make data available for calculating instrument drift over time. If only one of these
conditions is documented during each calibration event, it will be difficult to determine how well an
instrument is holding its calibration over long periods of time. Excessive drift is often an indicator
of impending failure, which is vital for any program of predictive maintenance or quality control.
Typically, the format for documenting both As-Found and As-Left data is a simple table showing
the points of calibration, the ideal instrument responses, the actual instrument responses, and the
calculated error at each point. The following table is an example for a pressure transmitter with a
range of 0 to 200 PSI over a five-point scale:
The following photograph shows a single-point “As-Found” calibration report on a temperature
indicating controller, showing the temperature of the calibration standard (−78.112 degrees Celsius),
the display of the instrument under test (IUT, −79 degrees Celsius), and the error between the two
(−0.888 degrees Celsius):
Note that the mathematical sign of the error is important. An instrument that registers −79
degrees when it should register −78.112 degrees exhibits a negative error, since its response is lower
(i.e. more negative) than it should be. Expressed mathematically: Error = IUT − Standard. When
the error must be expressed in percentage of span, the formula becomes:
18.3.3 Up-tests and Down-tests
It is not uncommon for calibration tables to show multiple calibration points going up as well as
going down, for the purpose of documenting hysteresis and deadband errors. Note the following
example, showing a transmitter with a maximum hysteresis of 0.313 % (the offending data points
are shown in bold-faced type):
Note again how error is expressed as either a positive or a negative quantity depending on whether
the instrument’s measured response is above or below what it should be under each condition. The
values of error appearing in this calibration table, expressed in percent of span, are all calculated by
the following formula:
In the course of performing such a directional calibration test, it is important not to overshoot
any of the test points. If you do happen to overshoot a test point in setting up one of the input
conditions for the instrument, simply “back up” the test stimulus and re-approach the test point
from the same direction as before. Unless each test point’s value is approached from the proper
direction, the data cannot be used to determine hysteresis/deadband error.
No comments:
Post a Comment